Skip to main content

Osmometer Fiske210

Advanced Instruments

Osmometer Fiske210

Single-Sample Osmometer

Model 210 Micro Osmometer determines the osmolality of solutions using freezing point depression (FPD). Ideal for serum, urine or other bodily fluids in clinical applications; and media or formulations in pharmaceutical, biotech or industrial applications.

Automated, single sample, designed to process a 20 mL sample, and 90-second test time. The Model 210 provides the best accuracy and precision in the market today.

Request pricingDescarregar catàleg

Product features:

  • 20 µL sample size is ideal for applications where the sample quantity is limited.
  • Unparalleled ease of use
  • Solid state cooling
  • Internal diagnostics
  • Ergonomic, compact design
  • On-board statistical analysis
  • Results recall up to the last 30 tests
  • Automated calibration

Technical specifications:

Specimen type Body fluids
Sample Volume 20 μL
Test Time 90 seconds
Sample Capacity Single Sample
Resolution 1 mOsm/kg H2O
Range 0 a 2000 mOsm/kg H2O
Linearity Less than ±1% from a straight line between O and 2000 mOsm/kg H20
Repeatability ±2 mOsm/kg Hp (1 S.D.) between O and 400 mOsm/kg H20 ;
±0.5% (1 S.D.) between 400 and 2000 mOsm/kg H20
Drift Less than 1 mOsm/kg H20 per month
Temperature Effects Less than 1 mOsm/kg H20 per 5ºC (9ºF) ambient temperature change
Communications  DTE RS-232 serial port, parallel printer port, and optional barcode scanner
Storage Temperature -40°C a+45°C
Electrical Voltage 100 a 230 VAC (50/60 Hz)
Power Consumption 60 Watts
Dimensions 30.5cm x 30.5cm x 24.1cm
Net Weight 6.2 kg
Shipping Weight 9.3 kg

Ask us any doubt and we will give you advice on buying, installation and maintainance

Model 210 Micro Osmometer determines the osmolality of solutions using freezing point depression (FPD). Ideal for serum, urine or other bodily fluids in clinical applications; and media or formulations in pharmaceutical, biotech or industrial applications.

Automated, single sample, designed to process a 20 mL sample, and 90-second test time. The Model 210 provides the best accuracy and precision in the market today.